Sensörler
|
Sensor Technology |
Michell Ceramic Moisture Sensor |
Sensor Version |
Easidew PRO I.S. |
Measurement Range |
-100 to +20°Cdp |
Calibration Range |
-100 to +20°Cdp
|
Calibration
|
Traceable to British (NPL) and American (NIST) National Humidity Standards |
Accuracy |
Dew point: ±1°C between -60 and +20°Cdp Moisture content: ±10% of reading Dew point: ±2°C between -60.1 and -100°Cdp
|
Resolution
|
0.1°C between +20 and -100°Cdp |
Analysis Pressure |
Up to 45 MPa (450 barg)
|
Operating Temperature
|
-40 to +60°C |
Sample Flow Rate |
1 to 5 l/min
|
Optional Pressure Sensor
|
0-138 barg (other ranges available) Accuracy: ±0.25% FS
|
Sertifikasyon
|
Hazardous Area Certification |
ATEX - II 1 G Ex ia IIC T4 Ga (-20°C = Ta = +70°C) FM - IS / I / 1 / ABCD / T4 Ta = +70°C CSA - IS Class I, Division 1, Groups ABCD T4 +70°C IECEx - Ex ia IIC T4 (-20°C = Ta = +70°C) GOST Ex |
Pattern Approval |
GOST-R, GOST-K |
Kontrol Ünitesi
|
Display |
Two line 6-digits LED, displaying moisture content / dew point (user toggle) and analysis pressure |
Analogue Output |
Two 4-20 mA (max load 500 Ω) User-configured for parameter, unit and range |
Digital Output |
RS485 Modbus RTU
|
Display Mode |
Moisture content (ppmV) Moisture content in natural gas (ppmV, lb/MMSCF, mg/m3) Dew point (°C or °F) Temperature (°C or °F) Pressure (psig, barg)
|
Pressure Compensation |
Fixed value (user-programmed) or dynamic input from optional pressure sensor
|
Display Resolution |
0.1°Cdp, 0.1°Fdp, 0.1 - 0.001 ppmV ideal gas (adjustable), 0.01 ppmV natural gas, 0.01 mg/m3, 0.01 lb/MMSCF, 1 psig, 0.1 barg
|
Alarms |
Four alarm relays. Control action and set-point are user-programmable Two Form C contacts rated 10 A, 240 V AC or 8 A, 24 V DC. Non-inductive load Two Form A contacts rated 5 A, 240 V AC or 4 A, 24 V DC. Non-inductive load
|
I.S. Barriers |
Galvanic isolation type, integrated to Control Unit
|
Power Supply |
90 to 264 V AC, 50/60Hz or 10 to 72 V DC, 10 W max power consumption
|
Interconnection Cable |
General instrument type, twisted pair, screened, single pair (two pairs with pressure sensor)
|
Enclosure |
19" sub rack unit Dimensions 132 x 483 x 256mm (h x w x d) (100mm minimum clearance depth for cables and vents)
|
Operating Environment |
Indoor, safe area, 0 to +50°C, < 90% RH
|
Birinci Sınıf Örnek Alma Sistemi
|
Enclosure |
304 stainless steel (EN1.4301) enclosure Option for complete enclosure in 316 stainless steel (EN1.4401) All fixtures stainless steel Galvanised steel internal mounting plate Open panel version available for indoor installation Dimensions 800 x 600 x 300mm (h x w x d) |
Enclosure Mounting |
Stainless steel wall mounting brackets |
Enclosure Ingress Protection |
IP66 |
Enclosure Temperature Control |
Heater/thermostat options for fixed set-point +20°C or adjustable set-point range 0 to control 50°C |
Heater Power Supply |
110/120 or 220/240/255 V AC, 47/63 Hz Power consumption 100 W max |
Operating Environment |